• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL CAVITRON JET PLUS ULTRASONIC SCALER AND AIR POLISHING PROPHYLAXIS SYSTEM; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL CAVITRON JET PLUS ULTRASONIC SCALER AND AIR POLISHING PROPHYLAXIS SYSTEM; SCALER, ULTRASONIC Back to Search Results
Model Number G137
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.Evaluation found no issues with unit heating inserts.Water pressure was at 22 psi with water flow over 100 cc/min at maximum setting.All electrical calibration points are within calibration.Operated unit for extended period with no heat development of insert.Cleaned up tubing, replaced water filter and depleted foot control batteries along with tuning/calibration of unit, all functions tested.The customer was advised to evaluate inserts for damage/wear and recommend denstply inserts only.
 
Event Description
While using a cavitron g137 scaler, the insert was heating up; no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON JET PLUS ULTRASONIC SCALER AND AIR POLISHING PROPHYLAXIS SYSTEM
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6194381
MDR Text Key63065112
Report Number2424472-2016-00096
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 12/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG137
Device Catalogue Number8187501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2016
Initial Date FDA Received12/21/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-