• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This event occurred in (b)(6).
 
Event Description
The customer received questionable elecsys ft4 ii assay results for three patient samples from a cobas 8000 e 602 module analyzer compared to an architect analyzer.The serial number was requested but was not provided.The results were provided as: (b)(6).The erroneous results were reported outside the laboratory.The patients were not adversely affected.
 
Manufacturer Narrative
Samples from the three patients were submitted for investigation and the customer's results for all three samples were reproduced.For the patient 1 and 3, an interfering factor was detected causing most likely the falsely elevated value of the ft4ii assay.This interference is documented in product labeling.For the patient 2, no interfering factor was detected.The difference in the ft4 values generated by the roche diagnostics method compared to the abbott method cannot be explained with available methods and the current state of the art.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6194383
MDR Text Key63068162
Report Number1823260-2016-02042
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot Number180429
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2016
Initial Date FDA Received12/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age57 YR
-
-