• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Gas Output Problem (1266)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 11/22/2016
Event Type  malfunction  
Manufacturer Narrative
Ge healthcare's investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported that, during a case, anesthetic agent output was noted to spike.The patient reportedly became hypotensive and was reportedly given a vasopressor to address.There was no reported patient injury.
 
Manufacturer Narrative
The hospital inspected the equipment and noted that the inflow/outflow/scavenging solenoid valve was free moving.Gehc recommended to the hospital to apply silicone to the mounting post to help prevent unwanted movement.The hospital inspected the equipment and noted that the inflow/outflow/scavenging solenoid valve was free moving.Gehc recommended to the hospital to apply silicone to the mounting post to help prevent unwanted movement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key6194500
MDR Text Key62976898
Report Number2112667-2016-02493
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042154
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2016
Initial Date FDA Received12/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/19/2017
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age69 YR
Patient Weight56
-
-