MEDTRONIC NAVIGATION, INC. (LOUISVILLE) NAV THORACIC PROBE TIP; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734680
Device Problems Bent (1059); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation of returned suspect probe finds that the reported problem was confirmed to be caused by physical damage of the instrument tip.Found tip of the probe was bent and twisted.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A medtronic representative reported that during a spinal fusion procedure the thoracic probe became bent and twisted.The medtronic representative reported the procedure was able to continue as planned.The surgeon completed the procedure with the use of the navigation system.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAV THORACIC PROBE TIP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027
7208902338
MDR Report Key6197056
MDR Text Key63067541
Report Number1723170-2016-05900
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00613994870063
UDI-Public00613994870063
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734680
Device Lot Number160830
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/29/2016
Initial Date FDA Received12/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-