Merge technical support instructed the user to reboot the hemo system.Device labeling, hemo v9.40 user manual, addresses such an occurrence with the following statement, "check that the crossover cable from the client to the hemo monitor is plugged in properly and lights are on.Reboot the hemo monitor pc.Exit and reopen the study.Answer yes to is patient still being monitored.Check that the client is properly configured to use the hemo monitor pc (in system config).If none of these work, contact technical support.".
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the hemo monitor was not communicating properly during a procedure where the patient had been sedated and active monitoring initiated.Subsequently, the hemo monitor and client pc were rebooted by the user during the procedure that resulted in a loss of patient monitoring.The delay was ~10-15 minutes while the hemo system rebooted.With merge hemo not capturing physiological data, there is a potential for delay in treatment that could result in harm to the patient.The customer reported that the procedure was completed successfully once the hemo system was rebooted.(b)(4).
|