MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 9.40.3
Device Problems Application Program Problem (2880); Blocked Connection (2888)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/23/2016
Event Type  malfunction  
Manufacturer Narrative
Merge technical support instructed the user to reboot the hemo system.Device labeling, hemo v9.40 user manual, addresses such an occurrence with the following statement, "check that the crossover cable from the client to the hemo monitor is plugged in properly and lights are on.Reboot the hemo monitor pc.Exit and reopen the study.Answer yes to is patient still being monitored.Check that the client is properly configured to use the hemo monitor pc (in system config).If none of these work, contact technical support.".
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the hemo monitor was not communicating properly during a procedure where the patient had been sedated and active monitoring initiated.Subsequently, the hemo monitor and client pc were rebooted by the user during the procedure that resulted in a loss of patient monitoring.The delay was ~10-15 minutes while the hemo system rebooted.With merge hemo not capturing physiological data, there is a potential for delay in treatment that could result in harm to the patient.The customer reported that the procedure was completed successfully once the hemo system was rebooted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key6197470
MDR Text Key63089307
Report Number2183926-2016-00821
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40.3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/23/2016
Initial Date FDA Received12/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-