Results: the distal tip of the cat6 was ovalized.The cat6 was also ovalized approximately 8.0 cm from the distal tip.The cat6 was fractured approximately 107.0 cm from the hub.Conclusions: evaluation of the returned device revealed that the distal tip of the cat6 was ovalized.This type of damage likely occurred due to forceful handling during use.Further evaluation of the returned device revealed that the cat6 was fractured.This damage was likely incidental and likely occurred due to forceful handling while packaging the device for return.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.
|