• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H749RB4225100
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).
 
Event Description
It was reported that a balloon rupture occurred.The target lesion was located in the moderately tortuous and severely calcified second right posterolateral (rpl) artery.A 10/2.5 flextome¿ cutting balloon¿was used for pre-dilation.It was noted that there was resistance encountered during insertion.The balloon was then initially inflated at 6atm and the 2nd inflation was at 8atm.During the third inflation, the balloon ruptured at 8 atmospheres.The procedure was completed with another of the same device.No patient complications were reported.
 
Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and inflation lumen which is evidence of a device leak.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located at the distal edge of the distal marker band.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip or blades.All blades were present and fully bonded to the balloon surface.The hypotube was kinked at several locations along its length.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
 
Event Description
It was reported that a balloon rupture occurred.The target lesion was located in the moderately tortuous and severely calcified second right posterolateral (rpl) artery.A 10/2.5 flextome¿ cutting balloon¿was used for pre-dilation.It was noted that there was resistance encountered during insertion.The balloon was then initially inflated at 6atm and the 2nd inflation was at 8atm.During the third inflation, the balloon ruptured at 8 atmospheres.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME¿ CUTTING BALLOON¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6199109
MDR Text Key63193035
Report Number2134265-2016-11966
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/08/2019
Device Model NumberH749RB4225100
Device Catalogue NumberRB422510
Device Lot Number0019009510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2016
Initial Date FDA Received12/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-