The complaint states upon inspection of trays, it was found that the spring on the spacer block sz5-6shim side was broken.The investigation confirmed the failure mode as reported.Care needs to be used when assembling these connections, as damage may occur resulting in the type of failure seen here.However, there is insufficient information supplied with this complaint to determine whether sufficient care was or was not taken.It should be noted that pra (b)(4) was released on (b)(6) 2014 in which the damage to the balseals was evaluated.The pra concluded that there is no patient harm as none of the failure modes have led to any patient harm.Furthermore, the assembly and disassembly of this device is carried out by the scrub nurse on a table away from the joint space, additionally the device cannot be disassembled within the joint space.It should also be noted that this issue will be further monitored in post market surveillance the complaint shall be closed with an undetermined conclusion and entered into the complaints system and monitored through trend analysis.If further information is received the complaint shall be reopened and investigated further.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|