• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 ATTUNE SPACER BLOCK; KNEE INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 ATTUNE SPACER BLOCK; KNEE INSTRUMENTS Back to Search Results
Catalog Number 254401014
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 12/13/2016
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).
 
Event Description
Upon inspection of trays, it was found that the spring on the spacer block sz5-6shim side was broken.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
The complaint states upon inspection of trays, it was found that the spring on the spacer block sz5-6shim side was broken.The investigation confirmed the failure mode as reported.Care needs to be used when assembling these connections, as damage may occur resulting in the type of failure seen here.However, there is insufficient information supplied with this complaint to determine whether sufficient care was or was not taken.It should be noted that pra (b)(4) was released on (b)(6) 2014 in which the damage to the balseals was evaluated.The pra concluded that there is no patient harm as none of the failure modes have led to any patient harm.Furthermore, the assembly and disassembly of this device is carried out by the scrub nurse on a table away from the joint space, additionally the device cannot be disassembled within the joint space.It should also be noted that this issue will be further monitored in post market surveillance the complaint shall be closed with an undetermined conclusion and entered into the complaints system and monitored through trend analysis.If further information is received the complaint shall be reopened and investigated further.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SPACER BLOCK
Type of Device
KNEE INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6200159
MDR Text Key63200767
Report Number1818910-2016-33798
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401014
Device Lot NumberBFA0P5A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2016
Initial Date FDA Received12/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/10/2017
02/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-