MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. LIMA HIP; REV BODY, STD. 90MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. LIMA HIP; REV BODY, STD. 90MM Back to Search Results
Catalog Number 428-00-090
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 11/30/2016
Event Type  Injury  
Event Description
Revision surgery - due to the previous stem breaking out of the posterior cortex post operation.
 
Manufacturer Narrative
The reason for this revision surgery was the patient's stem broke out of the posterior cortex, post operation.The in-vivo length of patient service for the implant was 1 day.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The root cause for the stem breaking through the posterior cortex was not reported.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIMA HIP
Type of Device
REV BODY, STD. 90MM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-5445
5128346255
MDR Report Key6201457
MDR Text Key63185442
Report Number1644408-2016-01039
Device Sequence Number1
Product Code LWJ
UDI-Device Identifier00888912081269
UDI-Public(01)00888912081269
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092331
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/20/2018
Device Catalogue Number428-00-090
Device Lot Number1305116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/01/2016
Initial Date FDA Received12/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
400-03-443, LOT 78B1071
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-