• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/22/2016
Event Type  Injury  
Manufacturer Narrative
Carefusion complaint number (b)(4).(b)(4).Results of investigation: a carefusion field service representative (fsr) evaluated the device onsite.The fsr checked the unit with the same settings as the customer reported and could not duplicate the reported issue.The fsr was informed by the customer that it is possible that there was a leak in the circuit that was being used at the time of the event and this could have caused the unit to not pressurize.The fsr evaluated the unit with a new circuit at various settings and no issued were found.The unit is working per manufacturer's specifications.
 
Event Description
The customer reported that the piston stopped and the customer could not get it to restart, while the device was in use on a patient.The patient was removed from the ventilator and placed on another unit.There was no patient harm.The customer changed the patient circuit and cap diaphragms, but the piston still would not start.The customer was able to perform the patient circuit calibration and performance check on the unit without any issues and was not able to duplicate the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6201473
MDR Text Key63171987
Report Number2021710-2016-05098
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2016
Initial Date FDA Received12/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-