Model Number 3100A |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/22/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).A carefusion field service representative (fsr) evaluated the device onsite.The ventilator was due for the 12k preventative maintenance.The fsr replaced the driver and installed the 12k preventative maintenance kit.The fsr installed a new circuit calibration label and a new 9 volt battery.The unit is operating to specification and is ready to be placed into service.The 3-ohm driver assembly was returned to the carefusion failure analysis laboratory.Upon completion of the device investigation, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the mean airway pressure (map) was set at 17 and was fluctuating between 15 and 20, while the device was in use on a patient.The customer replaced the circuit and changed out the cap diaphragms, but this did not resolve the issue.The patient was removed from the ventilator and placed on another unit.There was no patient harm.The customer reported evaluating the ventilator after the incident and could not duplicate the issue.The ventilator passed the patient circuit calibration and performance check and the map was steady.
|
|
Manufacturer Narrative
|
The vyaire medical failure analysis laboratory received the suspect component, a 3100a 3-ohm driver assembly, and evaluated the device.An evaluation of the component did not duplicate the reported issue.There were no fluctuations observed in the mean airway pressure (map).The map remained stable.
|
|
Search Alerts/Recalls
|