ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported that during calibration/setup of the system, upon placing the cassette, the screen turned white and the system froze.The case was initiated and completed as planned using an alternate system.There was no patient involvement.Additional information has been requested and received.
 
Manufacturer Narrative
The system was examined and the reported event(s) were not replicated.The random access memory (ram) was replaced as a preventive measure.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.Based on qa assessment, the product met specifications at the time of release.Ram was received for evaluation.A visual assessment of the returned sample revealed no obvious nonconformity.The ram was installed into a calibrated c system for functional testing.The system was found to meet relevant specifications the system was found to meet specification.Therefore, the root cause of the reported system issues cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6202044
MDR Text Key63521896
Report Number2028159-2016-05901
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2016
Initial Date FDA Received12/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-