• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event Description
It was reported a tether broke when attempting to remove the tether off the stent during a ureteroscopy with laser and basket stone removal.Reportedly the physician leaves the tether on the stent and then repositions the stent with the tether.The stent was reported to have already been repositioned when the breakage of the tether occurred and was removed by hand without instrumentation.Reportedly the stent was not removed and was reported to have been implanted on (b)(6) 2016.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Investigation - evaluation: a review of complaint history, device record history, manufacturing instructions, specifications, and quality control data was conducted during the investigation.The complaint device was not returned; therefore no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that would contribute to this failure mode.There were no other reported complaints for this lot number.Based on the information provided, no product returned, and the results of our investigation; a definitive root cause could not be determined.Measures have been conducted to address this failure mode.Monitoring will continue to be performed for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA FIRM URETERAL STENT SET
Type of Device
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
rita harden
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6202462
MDR Text Key63221531
Report Number1820334-2016-01546
Device Sequence Number1
Product Code FAD
UDI-Device Identifier10827002234069
UDI-Public(01)10827002234069(17)190428(10)6924772
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961446
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUFH-626-RT1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/01/2016
Initial Date FDA Received12/23/2016
Supplement Dates Manufacturer Received09/01/2017
Supplement Dates FDA Received09/06/2017
Date Device Manufactured04/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-