Model Number 801074 |
Device Problem
Out-Of-Box Failure (2311)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Lot #/model #: (21)00877.(b)(4).The defective o2 sensor was tested by the product surveillance technician.Oxygen percentages displayed on the ccm and measured with an external oxygen analyzer were within specifications compared to the set point.Nothing observed that would cause failure was noted during visual inspection.Testing evaluation included installing the returned o2 sensor into a lab use only epgs and connected the epgs to a system 1 simulator.Connected the epgs to oxygen and air, entered a perfusion screen on the ccm and waited the 15 minute warmup period.After the 15 minute warmup period, calibration of the epgs was initiated and passed.The measurement of the dc output voltage from the o2 sensor at 5 l/min and 100% o2 was 1.74 volts which is within the specification of.55-2.758 volts.Operated the epgs through its range of o2 percentages and measured the o2% output on the ccm and with an external o2 analyzer.The results are shown in the table below with no inaccuracies found.(b)(6).
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center the unit failed o2 sensor accuracy tests in phase 2 testing.It was reported that the o2 sensor failed out of the box.No patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) replaced the oxygen (o2) sensor.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Search Alerts/Recalls
|