DEPUY ORTHOPAEDICS, INC. PFC*CALIBRATED PAT CUT GDE; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 865034
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The saw capture was broken off while cutting the patella.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
Examination of the submitted cutting guide confirmed one of the two saw captures has disassembled from the device.The root cause is being attributed to device wear from normal use and servicing.Preliminary risk assessment (b)(4) was conducted.Based on the root cause determination of device wear from normal use and servicing, the need for corrective action is not indicated.Continue to monitor through (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6204554
MDR Text Key63260287
Report Number1818910-2016-33875
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number865034
Device Lot NumberSO2024242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2016
Initial Date FDA Received12/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/05/2017
01/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age63 YR
-
-