Model Number ERIS-CF25 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/24/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to olympus for evaluation.The cause of the reported event could not be determined; however, the most probable cause could be attributed to the operator's technique.The instruction manual states, ¿always keep sheaths parallel to one another when assembling and disassembling.If the inner sheath is inserted or removed at an angle to the outer sheath, the lateral force applied to the inner sheath may crack, loosen, break, or otherwise damage the sheath¿s insulated distal tip.A broken tip, or fragments of a damaged tip, can potentially pass through the outer sheath and into the patient.If drag or resistance is encountered during assembly or disassembly, stop¿align working element and sheath parallel to one another before proceeding.".
|
|
Event Description
|
Olympus received a medwatch report on december 8, 2016 stating that during a cystoscopy procedure, pieces of the resectoscope tip broke off inside the patient.The physician was able to recover all of the pieces using a grasper (model unknown) and saline was used to irrigate the patient.The procedure was completed.No patient injury.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.The evaluation confirmed the reported device damage.The device was received with the ceramic insulation tip completely broken off and missing from the distal end.The broken tip was not returned to olympus.The device was inspected and noted foreign material at the base of the shaft.There were dents and scratches also noted on the shaft.In addition, the red aligning dot was found missing from the release body section of the device.Based on the investigation findings and similar reported events, the cause of the reported event is likely due to the operator¿s technique.The instruction manual for use states, ¿do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged.Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip.If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use.¿.
|
|
Search Alerts/Recalls
|
|