Catalog Number 06437281190 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/15/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).(b)(4).
|
|
Event Description
|
The customer complained that the results for 9 patients tested for elecsys ft4 ii assay (ft4 ii) on the roche diagnostics elecsys e170 modular analytics immunoassay analyzer did not match the patient¿s clinical picture.Comparison tests were performed between the e170 analyzer and the equilibrium dialysis method and erroneous ft4 ii results were identified for 8 patients.The erroneous results were reported outside of the laboratory.Refer to the attached data for patient results.No adverse event occurred.The e170 analyzer serial number was (b)(4).
|
|
Manufacturer Narrative
|
A specific root cause was not identified.The patient samples were submitted for investigation.The ft4 ii results generated at the customer site were confirmed.Interference substance testing was performed and no interfering factors were identified.Possible factors to consider are that different technologies may generate mathematically different ft4 results, the reference range for the ft4 ii assay is different based on age and gender and dialysis testing methods can be affected by the absorption of thyroxine causing low ft4 results.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
|
|
Manufacturer Narrative
|
Based on the data provided, a general reagent issue can most likely be excluded.
|
|
Search Alerts/Recalls
|