Catalog Number 14324
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2016
Event Type  malfunction  
Manufacturer Narrative
Catalog number - the correct catalog number is 14324-97.(b)(4).
 
Event Description
A customer reported a positive result with a cyclesure 24 biological indicator (bi) after a completed sterrad 100s cycle.The bi was incubated for 24 hours.The chemical indicator (ci) changed color correctly.The affected load was not released.There was no report of infection, injury or harm to patient(s) associated with this issue.Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured.Therefore, as a matter of policy asp had decided to report all incidents of positive cyclesure 24 biological indicators.
 
Manufacturer Narrative
Asp investigation summary: the investigation included a review of the device history record (dhr), system risk analysis (sra), and product return and retains analysis.The dhr was reviewed and the involved lot met manufacturer specifications at the time of release.No anomalies were observed that would contribute to the customer's experienced issue.Trending analysis by lot number was reviewed from 07/27/2016 to 12/06/2016 and trending was not exceeded.The sra indicates the risk associated with hazard of 'quality problem with no impact on safety is "low." the single cyclesure® 24 bi was returned for visual inspection.The chemical indicator disc was gold in color which indicates exposure to hydrogen peroxide.There was yellow media remaining in the vial, confirming the suspect positive result.No manufacturing related anomalies observed.Thirty-two retains bis were subject to functional evaluation.All thirty-two bis met specification.It is unlikely the suspected positive bi was caused by a manufacturing issue as the retains met functional specification, dhr review found no anomalies that would contribute to a positive bi result and lot history review found trending was not exceeded in this lot.Also, no manufacturing anomalies were observed in the returned suspect bi that would contribute to a positive bi result.An issue with sterrad® performance is also unlikely as the cycle passed and the chemical indicator disc changed correctly.The customer stated subsequent bi was negative for growth.The cyclesure® retains met functional specification.As a result, there is not an issue with product performance when used per the instructions for use (ifu).The issue will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR
Type of Device
INDICATOR, BIOLOGICAL (FRC)
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
joaquin kurz
irvine, CA 92618
9497893837
MDR Report Key6209208
MDR Text Key63486595
Report Number2084725-2016-00794
Device Sequence Number1
Product Code FRC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K994055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2017
Device Catalogue Number14324
Device Lot Number20916112
Other Device ID Number14324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/04/2016
Initial Date FDA Received12/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/17/2017
03/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-