Report source: (b)(6).Complaint sample was evaluated and the reported event was confirmed.The device exhibits minor damage on the threads, specifically nicks in several locations.Regardless of the damage to the threads, the thread gage could be successfully threaded onto the rod and fully seated.The inserter exhibits signs of repeated use over a potential field service lifetime of approximately 4 years.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.Root cause was unable to be determined however; possible cause of the failure may be due to improper handling.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|