Model Number 3100B |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/29/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that after installing the patient circuit and connecting the supply gasses and powering up the unit, the oscillator started to run without pressurizing the circuit and pressing the start/stop switch.There was no patient involvement associated with the reported issue.
|
|
Manufacturer Narrative
|
The vyaire medical failure analysis laboratory received the suspect component, a 3100b driver displacement indicator (ddi) printed circuit board assembly (pcba), and evaluated the device.An evaluation of the component did not duplicate the reported issue as the component meets manufacturer's specifications.
|
|
Manufacturer Narrative
|
The vyaire medical factory service center and failure analysis laboratory received the suspect components, a 3100b 3 ohm driver assembly, a 3100b driver controller board, and a 3100b alarm printed circuit board assembly (pcba), and evaluated the devices.An evaluation of the components duplicated the reported issue and isolated the issue to pin 4 of ic u6 on the driver controller board was stuck in high state.
|
|