CAREFUSION, INC 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/29/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that after installing the patient circuit and connecting the supply gasses and powering up the unit, the oscillator started to run without pressurizing the circuit and pressing the start/stop switch.There was no patient involvement associated with the reported issue.
 
Manufacturer Narrative
The vyaire medical failure analysis laboratory received the suspect component, a 3100b driver displacement indicator (ddi) printed circuit board assembly (pcba), and evaluated the device.An evaluation of the component did not duplicate the reported issue as the component meets manufacturer's specifications.
 
Manufacturer Narrative
The vyaire medical factory service center and failure analysis laboratory received the suspect components, a 3100b 3 ohm driver assembly, a 3100b driver controller board, and a 3100b alarm printed circuit board assembly (pcba), and evaluated the devices.An evaluation of the components duplicated the reported issue and isolated the issue to pin 4 of ic u6 on the driver controller board was stuck in high state.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6210828
MDR Text Key63710747
Report Number2021710-2016-05115
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 11/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2016
Initial Date FDA Received12/28/2016
Supplement Dates Manufacturer Received08/02/2017
11/13/2017
Supplement Dates FDA Received08/09/2017
11/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-