• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2016
Event Type  malfunction  
Manufacturer Narrative
The exact age of the patient is unknown, however, it was reported the patient was over 18 years.(b)(4).A visual examination of the returned uphold¿ lite revealed that the suture on the blue dilator was broken.There was a tool mark on the suture.The dart was missing and was not returned.In addition, the suture on the blue with white stripe dilator was broken and the dart was missing and not returned.Analysis revealed that the catch of the capio suture capturing device was bent.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
 
Event Description
It reported to boston scientific corporation that an uphold¿ lite was used during an anterior repair of pelvic floor procedure on (b)(6) 2016.According to the complainant, during the procedure, the needle detached from the arm assembly during placement of the mesh and retrieved from the capio device.The procedure was completed with another uphold¿ lite device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6211875
MDR Text Key63478979
Report Number3005099803-2016-03901
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public(01)08714729839200(17)20190315(10)ML00003568
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date03/15/2019
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot NumberML00003568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2016
Initial Date FDA Received12/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-