BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The exact age of the patient is unknown, however, it was reported the patient was over 18 years.(b)(4).A visual examination of the returned uphold¿ lite revealed that the suture on the blue dilator was broken.There was a tool mark on the suture.The dart was missing and was not returned.In addition, the suture on the blue with white stripe dilator was broken and the dart was missing and not returned.Analysis revealed that the catch of the capio suture capturing device was bent.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
|
|
Event Description
|
It reported to boston scientific corporation that an uphold¿ lite was used during an anterior repair of pelvic floor procedure on (b)(6) 2016.According to the complainant, during the procedure, the needle detached from the arm assembly during placement of the mesh and retrieved from the capio device.The procedure was completed with another uphold¿ lite device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Search Alerts/Recalls
|
|
|