ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8 300; TRANSDUCER, PRESSURE,CATHETER TIP Back to Search Results
Model Number C12358
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of the investigation concluded the tip coil of the radiopaque tip and the corewire had been fractured at the distal tip.There was evidence of a weld joint at the distal end of the jacket.The distal sections of both the tip coil and corewire were not returned.The corewire had been bent.The device met specifications prior to leaving sjm manufacturing facilities as supported by a review of the device history record.There was no evidence found to suggest the incident was due to an intrinsic defect in the returned device.The cause of the guidewire damage is consistent with forcible contact during use.The pressurewire aeris instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire aeris instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire aeris instructions for use (ifu) states that the user should advance or withdraw the pressurewire slowly and never push, withdraw or torque the pressurewire if it meets resistance.
 
Event Description
When the pressurewire aeris was advancing down the vessel, the device broke into two pieces in the coronary artery.It was retrieved with no complications to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE WIRE AERIS G8 300
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6212117
MDR Text Key63474254
Report Number3008452825-2016-00170
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date05/31/2018
Device Model NumberC12358
Device Catalogue NumberC12358
Device Lot Number5555469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2016
Initial Date FDA Received12/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-