• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ELUNA 8 DR-T PROMRI; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ELUNA 8 DR-T PROMRI; PACEMAKER Back to Search Results
Model Number 394969
Device Problem Impedance Problem (2950)
Patient Problem Discomfort (2330)
Event Date 12/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Lead check failed for the ra lead and it flipped from bipolar to unipolar mode, twice since implant, and both times there was a slight decrease in impedance.All other values are within normal limits.A change in p-wave amplitude was noted when the lead check is failing, which indicates there could be an issue with the lead.Patient is symptomatic when the ra lead polarity switches to unipolar.The clinician requested disabling lead check.Clinician was advised that lead check is a safety feature for lead failure and disabling this feature removes that safety feature and then walked through to disable lead check.Clinician also resolved conflict of not being allowed to have capture control on and lead check off and also disabled atrial capture control.Clinician stated he would monitor patient and bring in again for further evaluation.
 
Manufacturer Narrative
As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUNA 8 DR-T PROMRI
Type of Device
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin D-123 59
GM  D-12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key6213858
MDR Text Key63560653
Report Number1028232-2016-05108
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number394969
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age61 YR
-
-