Lead check failed for the ra lead and it flipped from bipolar to unipolar mode, twice since implant, and both times there was a slight decrease in impedance.All other values are within normal limits.A change in p-wave amplitude was noted when the lead check is failing, which indicates there could be an issue with the lead.Patient is symptomatic when the ra lead polarity switches to unipolar.The clinician requested disabling lead check.Clinician was advised that lead check is a safety feature for lead failure and disabling this feature removes that safety feature and then walked through to disable lead check.Clinician also resolved conflict of not being allowed to have capture control on and lead check off and also disabled atrial capture control.Clinician stated he would monitor patient and bring in again for further evaluation.
|
As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
|