COVIDIEN KANGAROO EPUMP - NEW; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problems Leak/Splash (1354); Device Displays Incorrect Message (2591); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 12/29/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer has an issue with an epump.The customer states the unit displays an error alarm.However, upon triage, worn shaft of the leaky gearbox was found, making this complaint reportable.
 
Manufacturer Narrative
The unit was triaged and the customer¿s reported condition was confirmed.The root cause is due to worn shaft of the leaky gearbox all device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO EPUMP - NEW
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524151
MDR Report Key6214293
MDR Text Key63822555
Report Number3008361498-2016-00247
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number382400
Device Catalogue Number382400
Device Lot NumberC0724513
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/17/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer Received08/17/2016
Supplement Dates FDA Received09/05/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-