Product analysis was performed on both the explanted generator and portions of the explanted lead.During the explant surgery a portion of the patient's lead was cut into three segments.Each segment was evaluated for continuity and no discontinuities were identified.There were no functional anomalies seen with the lead.However only a portion of the lead was returned, no conclusions can be drawn regarding the portion of the lead that was not returned.Setscrew marks were observed on the connector pin of the lead and provide evidence of a proper mechanical contact between conductive surfaces of both the generator and connector pin, thereby ensuring a good electrical connection to the lead.The generator was also evaluated and no functional anomalies were identified with the generator.The data on the generator was reviewed and it was noted that there was a change >25% in impedance on the day following the explant surgery.No other relevant information has been received to date.
|