• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2016
Event Type  malfunction  
Manufacturer Narrative
The autopulse platform (s/n: (b)(4)) was returned to zoll for evaluation.The primary complaint of user advisory (ua) 45 (not at "home" position after power-on/restart) error message was confirmed during evaluation.The error message appeared upon powering on the platform.Review of the archive data confirmed that ua 45 error had occurred on the reported date of event.Based on the archive data, the drive shaft is not at the home position and this typical fault occurs when the user did not pull up the lifeband completely before powering on the platform.This user advisory will persist until the driveshaft is returned to its home position.Per the autopulse user guide instruct, to clear ua 45, the operator needs to pull up the lifeband until the chest bands are full extended.This action will move the driveshaft to its home position.Visual inspection was performed and found (unrelated to the reported complaint) cracks to the top, front, and bottom covers.The autopulse platform is a reusable device and was manufactured on 09/14/2011.Therefore, this type of physical damages is characteristic of normal wear and tear for the life of the device.The damaged parts observed during visual inspection were replaced.To resolve the issue, the driveshaft was returned to the home position.The autopulse was functionally tested and passed without errors.The platform passed all final testing criteria.Device history record (dhr) was reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
It was reported that the autopulse platform (s/n: (b)(4)) displayed a user advisory (ua) 45 (not at "home" position after power-on/restart) error message.According to the reporter, attempts were made to resolve the issue, however, the problem continued.There is no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key6214382
MDR Text Key63624625
Report Number3010617000-2016-00952
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 12/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/07/2016
Initial Date FDA Received12/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-