• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS Back to Search Results
Model Number SIPAP
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/16/2016
Event Type  malfunction  
Manufacturer Narrative
Carefusion complaint number (b)(4).(b)(4).At this time, carefusion has not received the suspect device/component from the customer for evaluation.At this time, the customer has not responded to requests for additional information regarding this event.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that when the unit was switched off, there was no audible alarm.As the customer reported that they were turning the unit off, there is no patient involvement nor adverse event.
 
Manufacturer Narrative
Results of investigation: the carefusion failure analysis laboratory received the suspect device, a sipap ventilator, and evaluated the device.An evaluation of the device duplicated the reported issue and isolated the issue to a super capacitor on the main printed circuit board assembly (pcba) not holding enough charge to activate the buzzer to create an audible tone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Type of Device
VENTILATOR, NON-CONTINUOUS
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6214713
MDR Text Key63733114
Report Number2021710-2016-05145
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K031745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIPAP
Device Catalogue Number27476-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-