SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number 518-062
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Unspecified Infection (1930); Injury (2348); Pericardial Effusion (3271)
Event Date 12/07/2016
Event Type  Injury  
Event Description
This procedure was to extract two cardiac leads due to cied system/pocket infection, indwelling for 123 months.The leads were prepped with spectranetics lead locking devices for traction.Upon removal of the rv lead a cardiac effusion was noted on tee.A sternotomy revealed a tear in the apex of the rv.The injury was repaired and the patient survived the procedure.This report is being made against the spectranetics lld as the traction platform.
 
Manufacturer Narrative
Follow up provided to supply partial udi, patient and device codes.
 
Manufacturer Narrative
Device 510k number has been corrected to reflect the most current and up to date number, as of the date of the initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal dr.
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key6214720
MDR Text Key63609322
Report Number1721279-2016-00198
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer ReceivedNot provided
07/24/2017
Supplement Dates FDA Received02/07/2017
08/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARDIAC LEAD, SJM 1642T (IMPL 123 MOS); CARDIAC LEAD, SJM 1646T (IMPL 123 MOS); SPECTRANETICS CVX-300 LASER SYSTEM; SPECTRANETICS GLIDELIGHT LASER SHEATH; SPECTRANETICS LEAD LOCKING DEVICES; SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age57 YR
Patient Weight107
-
-