• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2016
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).Lens not returned.
 
Event Description
The reporter indicated that the surgeon was attempting to implant a 13.2 mm vticm5_13.2 implantable collamer lens, -12/1.5/103 diopter in the patient's right eye (od) and it was noted that the lens was torn.
 
Manufacturer Narrative
Product evaluation found that the lens was returned in liquid and in the lens case/vial.Visual inspection found the lens haptic broken.No similar complaints were reported for units within the same lot.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key6214842
MDR Text Key63613864
Report Number2023826-2016-01846
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date04/30/2019
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2016
Initial Date FDA Received12/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age33 YR
-
-