ORIGEN BIOMEDICAL INC. ECMO DOUBLE LUMEN VV CANNULA 28 F; CANNULA, CATHETER Back to Search Results
Catalog Number VV28F
Device Problems Disconnection (1171); Physical Property Issue (3008)
Patient Problem Blood Loss (2597)
Event Date 11/23/2016
Event Type  malfunction  
Event Description
Patient was on vv ecmo, venous side of double lumen cannula came disconnected from company manufactured y site causing blood loss and temporary disconnection from ecmo circuit.Patient with additional venous cannula allowing maintaining of some ecmo support while preparing to change out double lumen cannula.Patient again required removal of ecmo support while replacing damaged equipment.Breakage of equipment caused patient to be removed from ecmo x3, blood loss that required additional 2 units of prbc's during and after event, required patient to be reintubated, exposed patient to possible contamination by having disconnection/breakage of ecmo cannula and exposing open central access to non-sterile environment.The patient appears to have returned to baseline.As far as what happened, our speculation is that the manufactured glue failed at the y connection.The patient had the cannula in place for several weeks.The patient had a second cannula in her groin area that did not pose any concerns or issues.
 
Event Description
Patient was on vv ecmo, venous side of double lumen cannula came disconnected from company manufactured y site causing blood loss and temporary disconnection from ecmo circuit.Patient with additional venous cannula allowing maintaining of some ecmo support while preparing to change out double lumen cannula.Patient again required removal of ecmo support while replacing damaged equipment.Breakage of equipment caused patient to be removed from ecmo x3, blood loss that required additional 2 units of prbc's during and after event, required patient to be reintubated, exposed patient to possible contamination by having disconnection/breakage of ecmo cannula and exposing open central access to non-sterile environment.The patient appears to have returned to baseline.As far as what happened, our speculation is that the manufactured glue failed at the y connection.The patient had the cannula in place for several weeks.The patient had a second cannula in her groin area that did not pose any concerns or issues.
 
Event Description
Patient was on vv ecmo, venous side of double lumen cannula came disconnected from company manufactured y site causing blood loss and temporary disconnection from ecmo circuit.Patient with additional venous cannula allowing maintaining of some ecmo support while preparing to change out double lumen cannula.Patient again required removal of ecmo support while replacing damaged equipment.Breakage of equipment caused patient to be removed from ecmo x3, blood loss that required additional 2 units of prbc's during and after event, required patient to be reintubated, exposed patient to possible contamination by having disconnection/breakage of ecmo cannula and exposing open central access to non-sterile environment.The patient appears to have returned to baseline.As far as what happened, our speculation is that the manufactured glue failed at the y connection.The patient had the cannula in place for several weeks.The patient had a second cannula in her groin area that did not pose any concerns or issues.The facility has maintained this product and have made it available for inspection to the manufacturer.We have shared details regarding the event and what happened to the patient.We shared detailed pictures that they (manufacturer) requested of the external and internal aspects of the cannula pieces.They in turn updated us on their failure testing and that they were able to replicate the event by placing a clamp on the clear section of tubing very close to the hub that seemed to create a stress fracture in the adhesive that predisposed weakness leading to failure.(we are not aware that a clamp was related to the event we experienced.) they shared that this cannula was among a lot from several years ago and that no one else has reported this event.They felt it was an isolated event and although the failure potential is present, unlikely to recur.They have created a technical bulletin that we were told would go to all customers who have ordered this product in the past few years and also with future product purchases.They verbalized plans to redesign the adhesive bonding process to address this issue.The company has been responsive in an effect to evaluate this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECMO DOUBLE LUMEN VV CANNULA 28 F
Type of Device
CANNULA, CATHETER
Manufacturer (Section D)
ORIGEN BIOMEDICAL INC.
7000 burleson rd.
bldg. d
austin TX 78744
MDR Report Key6215664
MDR Text Key63645842
Report Number6215664
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/29/2016,12/07/2016
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date08/30/2018
Device Catalogue NumberVV28F
Device Lot NumberN18487
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/29/2016
Device Age14 DY
Event Location Hospital
Date Report to Manufacturer11/29/2016
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/30/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/05/2017
01/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
UNKNOWN.; UNKNOWN.; UNKNOWN.
Patient Age7 YR
Patient Weight50
-
-