Product analysis was completed on the returned generator.Visual examination showed only observations consistent with the explant process.No other surface abnormalities were noted on this device or the internal printed circuit board.Review of the internal data from the pulse generator showed indications of increased impedance during device implant.On (b)(6) 2015, the impedance changed from a pre-change value of 7561 ohms to a post-change value of 9924 ohms.The device was returned in a pulse disabled condition that was due in part to the increased impedance which required a higher output voltage.The pulse disabled status of the generator would not reset.With the pulse generator case removed and the battery still attached to the pcba, the battery measured 1.762 volts, confirming an end of service battery condition.The internal device data revealed that 85.109% of the battery had been consumed.Electrical tests results show that the pcba performed as expected and passed all required quality tests.Other than the evidence of high impedance and its contribution to the end of service of the generator, there were no additional performance or any other type of adverse conditions found with the pulse generator.No additional pertinent information has been received to date.
|