COOK INC THREE-WAY PLASTIC STOPCOCK; KGZ ACCESSORIES, CATHETER Back to Search Results
Model Number N/A
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event Description
It was reported a high pressure 3 three-way stopcock being utilized for chemo embolization of the lung leaked outside of the syringe.The procedure was completed using another device.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.
 
Manufacturer Narrative
A review of the complaint history, drawings, device history record, manufacturing instructions, quality control, and visual inspection of the returned device was conducted during the investigation.One three-way plastic stopcock was returned for evaluation.Stopcock leakage was observed during testing.Also, a 2mm crack was found in the base of the device.Instructions are in place to verify that the device is manufactured to specification.There is no evidence to suggest the product was not made to specifications.Review of the device history record shows no nonconforming events that would contribute to this failure mode.There were no other reported complaints for this lot number.Based on the information provided and the results of our investigation, a definitive root cause could not be determined.Measures have been conducted to address this failure mode.Monitoring will continue to be performed for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE-WAY PLASTIC STOPCOCK
Type of Device
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6218604
MDR Text Key63742426
Report Number1820334-2017-00024
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier00827002002197
UDI-Public(01)00827002002197(17)190601(10)5083419
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberPTWS-2FLL-MLL-R
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2016
Initial Date FDA Received01/03/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-