• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEWPORT MEDICAL INSTRUMENTS, INC HT70 PLUS VENTILATOR; CONTINUOUS VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEWPORT MEDICAL INSTRUMENTS, INC HT70 PLUS VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number HT70 PLUS
Device Problem Sticking (1597)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the ht70 plus ventilator gets stuck on the six picture screen and does not go to the normal ventilation screen.Patient involvement is unknown.
 
Manufacturer Narrative
(b)(4).The covidien service engineer (se) inspected the device and verified the reported issue.The se replaced the single board computer (sbc) printed circuit board (pcb).The se performed extended self-testing on the device and all tests passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT70 PLUS VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENTS, INC
1620 sunflower ave.
costa mesa CA 92626
Manufacturer (Section G)
NEWPORT MEDICAL INSTRUMENTS, INC
1620 sunflower ave.
costa mesa CA 92626
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key6219439
MDR Text Key64031565
Report Number2023050-2017-00006
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHT70 PLUS
Device Catalogue NumberHT70PM-WW-NA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/06/2016
Initial Date FDA Received01/03/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-