CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Under-Sensing (1661)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2016
Event Type  malfunction  
Event Description
It was reported that heart rate could not be detected on the patient¿s generator.The output currents were reportedly left on during this attempt to detect heart rate.The generator device history record was reviewed and showed all specifications were met prior to distribution.The trim test tab operation record showed that the generator is a mechanically routed device.No additional pertinent information has been received to date.
 
Event Description
Programming history and internal device data from the generator was reviewed.In the last available appointment, the data shows three lines of interrogation or diagnostic data where tachycardia detection was enabled.The foreground heart rate measured 80.3 for those three lines.The sensitivity value used in these lines was setting 5.The internal data showed 89 tachycardia detections and 76 delivered autostimulations.Tachycardia detection and output currents were both on throughout the office visit.No anomalies were apparent from the data present.No additional pertinent information has been received to date.
 
Event Description
It was reported that the heart rate detection feature on the patient¿s generator was functioning as intended.No additional pertinent information has been received to date.
 
Event Description
The patient's generator was prophylactically replaced and the device was returned to the manufacturer for analysis.The generator has been received, but analysis has not been completed to date.
 
Manufacturer Narrative
(b)(4).
 
Event Description
Analysis on the generator was approved.The device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.There were no performance or any other type of adverse condition found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6219564
MDR Text Key63761267
Report Number1644487-2017-02966
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/10/2017
Device Model Number106
Device Lot Number203513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2017
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 12/07/2016
Initial Date FDA Received01/03/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
07/21/2017
08/22/2017
Supplement Dates FDA Received01/27/2017
03/26/2017
08/03/2017
08/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age43 YR
-
-