DEPUY SYNTHES POWER TOOLS STANDARD CRANIOTOME, BLACK MAX; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - CRANIOTOME Back to Search Results
Catalog Number B-GREEN
Device Problems Break (1069); Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Gtin is unavailable as the product made prior to compliance date; (b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that it was observed that the craniotome device had excessive temperature and breakage of parts.During service and evaluation, it was observed that the device neuro tip was damaged.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD CRANIOTOME, BLACK MAX
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY - CRANIOTOME
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6219775
MDR Text Key63771877
Report Number1045834-2017-10003
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K974025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB-GREEN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2016
Initial Date Manufacturer Received 12/09/2016
Initial Date FDA Received01/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-