Results: the cat6 was kinked in multiple locations along its distal shaft and ovalized approximately 1.0 cm from its distal tip.Conclusions: evaluation of the returned device revealed that the cat6 was ovalized near its distal end.If the peelable sheath is not used during insertion and the cat6 is forcefully gripped, the distal shaft may become ovalized.Further evaluation revealed that the cat6 was kinked in multiple locations along its distal shaft.This damage likely occurred due to forceful manipulation of the already-ovalized cat6 during insertion into the sheath.The non-penumbra sheath mentioned in the complaint was not returned for evaluation.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure using an indigo system aspiration catheter 6 (cat6).During the procedure, while inserting the cat6 into a non-penumbra sheath, the physician did not use the peelable sheath and experienced resistance.Consequently, the tip of the cat6 became kinked and was removed.The procedure was then completed using a new cat6 and the same non-penumbra sheath.There was no report of an adverse effect to the patient.
|