Following the event a steris technician arrived at the facility, inspected the reliance vision washer/disinfector and found damage on the cable door assembly.The cables were bare and created shorts in the electrical circuit, which shut the washer down.The shorts also caused the solenoid valve and door switch to fail.The damage was caused by the improper orientation of the cable door assembly.The dhr was reviewed and no issues were noted.The cable door assembly is tested during manufacturing for proper operation and again during final testing.No issues were noted during manufacturing and final testing.When the washer was installed no issues were noted with the cable door assembly.During the steris technician's inspection of the washer he was unable to determine the cause of the improper orientation of the cable door assembly.The technician replaced the cable door assembly, door switch, pneumatic valve, and ran a full test cycle.The unit was found operational and returned to service.No additional issues have been reported.
|