• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES SELZACH PATIENT SPECIFIC SPLINT ORTHOGNATHIC, INTERMEDIATE; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES SELZACH PATIENT SPECIFIC SPLINT ORTHOGNATHIC, INTERMEDIATE; TEMPLATE Back to Search Results
Catalog Number SD900.105
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device is an instrument and is not implanted / explanted.Device is not expected to be returned for manufacturer review/investigation.(b)(6).Device history records review was completed for part# sd900.105, lot# me16ocosov.Manufacturing location: (b)(4), manufacturing date: nov 24, 2016.No non conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follow: it was reported that on (b)(6) 2016, patient underwent the surgery for a bimaxillary osteotomy (with genioplasty).Surgeon was using proplan splints, with maxilla movement first.The intermediate splint was placed and did not fit.This caused an approximate 15 minutes delay to the surgery while the cause of the custom splint not fitting was being determined.Surgeon was concerned the mandible movements are not accurate.The maxillary plate was perfect with cutting guides.There was no adverse event to the patient as a result of this incident.The surgeon was pleased with outcome after the final result due to custom maxillary plate.Splint was not used for this portion of the procedure.This report is for one (1) patient specific splint orthognathic intermediate.This is report 1 of 1 for com-(b)(4).
 
Manufacturer Narrative
Additional narrative: a manufacturing investigation was performed.The received information was forwarded to the responsible supplier for investigation: pars not received for investigation.For this complaint, the device history record was investigated for the intermediate splint.No inaccuracies were found on the image quality, segmentation, planning and guide design steps.Some minor inaccuracies were found between the registration of the dental cast and the provided ct images of the patient.These could have been caused either by the age of the dental cast being older than the ct images provided (ct scan date is from (b)(6) 2016; date of the impression of the dental cast is unknown, however these were provided to materialise on (b)(6) 2016) and/or by any orthodontic treatment that may have been ongoing on the patient in between the scan date and the surgery time (surgery date on (b)(6) 2016).These discrepancies between the ct scans and the dental cast were noted during the pre-planning phase and were discussed during the planning session; in consequence the surgeon was made aware of potential interferences with the splint.Based on the available information, the impact of these differences in the performance of the intermediate splint cannot be confirmed nor excluded, and therefore no root cause could be found to explain the alleged issue by the complainant.Since no root cause could be found to explain the alleged issue by the complainant; no actions would be defined for this complaint.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT SPECIFIC SPLINT ORTHOGNATHIC, INTERMEDIATE
Type of Device
TEMPLATE
Manufacturer (Section D)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ  CH2545
Manufacturer (Section G)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ   CH2545
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6224821
MDR Text Key63962405
Report Number3000270450-2017-10001
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberSD900.105
Device Lot NumberME16OCOSOV
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/10/2016
Initial Date FDA Received01/05/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-