MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during a procedure, the system stopped.Additional information has been requested.
 
Manufacturer Narrative
(b)(4).
 
Event Description
Additional information was received from the operating room technician, who reported a system advisory displayed during surgery.
 
Manufacturer Narrative
The system was examined and the reported ¿shut down¿ was not replicated.However, the event log review showed several instances of the same reported system messages (sms).The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on september 1, 2016.Based on qa assessment, the product met specifications at the time of release.It remains inconclusive whether the system actually ¿shut down¿ or the customer was referring to this ¿safe state¿ that the system reverts to when critical fluidics functionality is affected.The company representative found no problems with the system.It should be noted that these instances are usually user-related issues, but can be addressed when reconfiguring the system settings.However, there is no evidence to suggest that customer involved (in this particular reported event) used the system incorrectly.The system was found to meet specifications.Therefore, the root cause of the reported event cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6227320
MDR Text Key64287758
Report Number2028159-2017-00086
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 05/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/15/2016
Initial Date FDA Received01/05/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received04/11/2017
05/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-