Case- (b)(4).The lot with nonconforming devices, 65863, was manufactured on 6/21/2016 prior to the implementation of the inspection or the welding procedure.The malfunction was found during a 100% inspection of devices that were returned from japan as part of a volunteer inventory swap.
|
Rework routers ((b)(4)) were used to open and inspect all five lots of emt1s that were returned from cmi per (b)(4).The inspections were completed on december 10, 2016.(b)(4) identified seven devices that did not meet the sort criteria, one of which had a dislodged proximal jaw drive subassembly pin.The pin dislodged during unpacking.This is documented on (b)(4).No patient involvement.
|