• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 60-03-00
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2016
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump console.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).A sorin group field service representative was dispatched to the facility to investigate.The service representative was informed by the user that the failure did not stop the pump.The service representative tested the device and confirmed the reported issue.Traces of liquids were identified in the external frame of the display.The touchscreen was replaced to resolve the issue and subsequent testing did not identify further issues.The unit was returned to service.The replaced touchscreen was returned to sorin (b)(4) for further investigation.Evaluation of the returned device confirmed the presence of fluid (mix of blood and saline water) located between the front plane and the touchscreen of the display.Additionally, a scratch was identified in the conductive coating of the display.The penetration of fluid into the internal layers of the display in combination with the damage of its conductive coating were determined to be the root cause of the reported issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.
 
Event Description
Sorin group (b)(4) received a report that the touch screen of the sorin centrifugal pump console stopped displaying text and symbols during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP CONSOLE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6235653
MDR Text Key64241337
Report Number9611109-2016-00935
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K020571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-03-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2016
Initial Date FDA Received01/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-