Same case as mdr id 2134265-2016-12207.It was reported that rotawire fracture occurred and the burr had a hole.The target lesion was located in the left anterior descending artery(lad).A 1.50 mm rotalink¿ burr and a 330 cm rotawire¿ were selected for use.During the procedure, while on dynaglide mode, the physician observed a hole in the shaft and rotaglide was leaking along with the drugs and other mixtures.The burr was then removed from the patient's body and replaced the burr with another of the same burr.Subsequently, the physician then noted that the distal end of the rotawire became detached.Snaring was done to retrieve the fragment; however, about 20 mm of the rotawire was left in the distal lad around the apex.The procedure was completed with another of the same burr.Stenting was done thereafter.There were no further patient complications reported and the patient's condition was fine.
|