|
Model Number 505 |
Device Problem
Improper Flow or Infusion (2954)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/20/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has been received for analysis, but the analysis has not been concluded.
|
|
Event Description
|
Medtronic received information that prior to the implant of this mechanical aortic valve, the valve was inspected and no abnormalities were found.Immediately following implant of this valve, an ultrasound showed that the patient¿s hemodynamics were not ideal.The physician explanted and replaced this valve with another valve of larger size.The operation was successfully and no patient injury was reported as a result of the event.Following implant, the patient's condition improved.
|
|
Manufacturer Narrative
|
Product analysis: upon receipt at medtronic¿s quality laboratory, the returned product specimen was visually examined.The valve appeared slightly discolored, showing evidence of blood contact.The valve was intact with no evidence of visible damage.The valve tissue annulus diameter was verified and met the specification.Both leaflets were in the closed position and appeared intact with no evidence of damage such as cracks and/or surface anomalies.Both inflow and outflow valve hinge mechanisms as well as the inflow and outflow orifices appeared to be intact with no evidence of damage.Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Based on the reported information, the cause of the event was a sizing issue which could potentially have been due to technical error.The device is considered acceptable.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|