The device was returned with no dispenser coil or introducer.The pusher coil was noted to be completely broken from the transition zone.The separated hypotube, along with the remainder proximal section, was not returned for evaluation.On the distal side, part of the implant coil was noted to be stretched out from the proximal tie knot location, and completely broken.The remaining implant coil along with the marker band was still attached to the pusher, with good tension in the monofilament.The distal black pet covering the heater coil had no signs of thermal damage.The returned product evaluation confirmed a broken coil.The reported event details indicated that the coil became detached during repositioning.It is possible that the device was subjected to forces that exceeded its tensile strength specifications; however, the exact root cause cannot be determined.The device was used during the same procedure as was reported in mfr report# 2032493-2017-00050.
|