Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: returned product consisted of a ffr comet pressure wire in one piece.The tip, device shaft and the sensor port were examined for damage or any irregularities.The tip showed no issues.The shaft showed one kink/bend approximately 124.5 cm from the tip.There was also some peeling of the coating approximately 125 cm from the tip.The sensor port was clear of any material.The complaint of a kink in the device was confirmed.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
|
Reportable based on device analysis completed on 15dec2016.It was reported that the base of the pressure wire kinked.The lesion was located in the moderately left anterior descending artery (lad).During the procedure, when the physician was about to insert the comet pressure wire into the y connector, the base part got kinked.There were no patient complications and the procedure was completed with another of the same device.Device analysis revealed some peeling of the coating approximately 125 cm from the tip.
|