• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number A2-80070
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem Not Applicable (3189)
Event Date 01/02/2013
Event Type  malfunction  
Manufacturer Narrative
User reported issue was confirmed.Corrective and preventive action (capa (b)(4)) had been issued to address this issue.The complaint was determined as not mdr reportable at the initial determination by following company procedures.This mdr is retrospectively filed as a corrective action to address one of the fda inspection observations made on 08/11/2016.
 
Event Description
The user reported that "leaking at drip from vented port" for the iv set.No patient was involved in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Type of Device
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
Manufacturer (Section G)
ZYNO MEDICAL, LLC
177 pine street
natick MA 01760
Manufacturer Contact
chaoyoung lee
177 pine street
natick, MA 01760
5086502008
MDR Report Key6247564
MDR Text Key64736220
Report Number3006575795-2016-00124
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00814371020051
UDI-Public00814371020051
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132841
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA2-80070
Device Lot Number55034KS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2013
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/07/2013
Initial Date FDA Received01/12/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-