MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Device Alarm System (1012); Device Stops Intermittently (1599); Invalid Sensing (2293); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 12/14/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information could not be obtained due to country privacy laws.(b)(4).The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported that, the scavenging sensor in various operating room showed different values.In urology room, the alarm activates intermittently at 45 lt/min, while in surgery room, the alarm is permanently active at 55 lt/min.There was no reported patient injury.
 
Manufacturer Narrative
The ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The software was updated to service pack 04 to resolve the reported issue.
 
Manufacturer Narrative
Ge healthcare product engineering reviewed this event and determined that this issue will not impact the machine function and ventilation will not be impacted.Thus, this case has been assessed as not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi
CH 
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 
MDR Report Key6247686
MDR Text Key65090206
Report Number9710602-2017-00005
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 01/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE H10
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/14/2016
Initial Date FDA Received01/12/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/14/2017
04/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-