• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. RECLAIM DIST REAMER EXTENSION; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. RECLAIM DIST REAMER EXTENSION; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 297500500
Device Problems Disassembly (1168); Mechanical Problem (1384); Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The white plastic handle was missing a flexible metal locking ring; therefore, it would not stay on the reamer handle.This was discovered before use on a patient.
 
Manufacturer Narrative
Additional narrative: this complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
Examination of the returned instrument confirmed the retaining spring component of the sleeve is missing; likely occurred during disassembly for the cleaning and sterilization process.The spring was not returned for evaluation.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DIST REAMER EXTENSION
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6249107
MDR Text Key64760255
Report Number1818910-2017-10826
Device Sequence Number1
Product Code HTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 02/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number297500500
Device Lot NumberNB99125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2017
Initial Date FDA Received01/13/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received02/09/2017
03/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-