The coil was returned without the detached distal coil segment.The microcatheter was not returned; therefore, testing could not be performed to evaluate the dimensional compatibility of the device with the coil.The pusher wire was noted to be kinked in several locations along the length, including the gold connector.Lead wire separation was also noted proximal to the warning mark.The body coil and pet were smashed just distal to the heater coil, twisting and damaging the lead wires.The solder joints appeared intact.The monofilament rebound was measured to be 0.175".The monofilament has a tail end that measures 0.006".The device exhibited evidence of excessive stress to the monofilament.Based on the evaluation of the returned coil segment and the information provided, the event most likely occurred as a result of multiple coil manipulations during the attempts to advance the coil through the microcatheter.
|
It was reported that during the procedure, an embolization coil would no longer advance in the microcatheter.The coil and the microcatheter were removed together as a single unit from the patient.After the devices were removed, the coil was noted to have detached and the distal coil segment still remained in the vessel.Another microcatheter and a solitaire stent coil retriever were used to remove the coil segment from the patient.There were no reported clinical sequelae.The patient's current status is reported to be "fine.".
|