• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problems Thermal Decomposition of Device (1071); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/16/2014
Event Type  Injury  
Manufacturer Narrative
(b)(4).Results of investigation: the carefusion failure analysis laboratory received the suspect component, a 3100b driver assembly, and evaluated the device.An evaluation of the component duplicated the reported issue and isolated the issue to a pole piece being out of place which caused the coil assembly to become damaged and burned.
 
Event Description
The customer reported that the mean airway pressure dropped and caused the driver on the unit to stop while the unit was in use on a patient.The ventilator alarmed and the user hand-ventilated the patient while troubleshooting the issue.There was no patient compromise.The user found the red cap diaphragm loose and replaced it.The user quickly re-pressurized the unit and restarted the ventilator.After a few minutes, the user started to hear a high pitched "squeak" with every beat of the ventilator.The user tried to ignore the "squeaking" as the settings were stable.After a few minutes, the user smelled an "electrical burning smell." the nurses felt the smell was coming from the back of the ventilator.There was no indication of any problems with the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6254392
MDR Text Key64916689
Report Number2021710-2017-05248
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 01/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number773967-RNT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2014
Initial Date FDA Received01/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-